The FDA (US Food and Drug Administration) regulatory framework for medical devices consists of multiple parts of 21 CFR (Code of Federal Regulations) Parts 800–898 (Subchapter H). These regulations cover the entire life cycle from device development and manufacturing to post-market surveillance. The chart below gives an overview of the core Parts of medical device regulation:
How the 21 CFR Parts connect and cover the entire life cycle
Every stage of a medical device’s journey, from development to withdrawal from the market, is governed by a corresponding Part:
Example compliance route: a Class II in vitro diagnostic (IVD) reagent
Taking a manufacturer of a Class II in vitro diagnostic reagent as an example, its compliance process must cover the following key steps:
-
Registration and listing: complete establishment registration and device listing under Part 807.
-
Pre-market submission: submit a 510(k) under Part 807 demonstrating that the product is substantially equivalent to a legally marketed predicate device.
-
Quality management: establish and operate a quality management system in accordance with Part 820 to ensure the production process complies with requirements.
-
Labelling and traceability: apply the UDI as required by Part 830 and upload the information to the GUDID database in a timely manner.
-
Post-market surveillance: report adverse events such as deviations in test results under Part 803 and initiate CAPA (corrective and preventive action) at the same time; where the FDA orders postmarket surveillance under Part 822, conduct the PMS study under Part 822 to continuously track product risks; and if the device is subject to tracking requirements, establish a device tracking system under Part 821