US Medical Device Registration Overview: This page covers the regulatory authority, device classification, registration process, local agent requirements and FAQs for medical device registration in the United States. Medical devices in the US are regulated by the FDA’s Center for Devices and Radiological Health (CDRH), and EVERBIZ can support you through the entire process from classification to clearance.
1. Regulatory Framework
Medical devices in the US are regulated by the Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration (FDA). The US is the world’s largest, most mature and most influential medical device market, and many other countries use FDA approval as a basis for expedited review.
Core legislation:
-
Federal Food, Drug, and Cosmetic Act (FD&C Act) – the principal statute
-
21 CFR Parts 800–1299 – medical device regulations (key parts include 800, 803, 806, 807, 820 and 830)
-
MDUFA V (Medical Device User Fee Amendments passed in 2022, covering FY2023–2027)
Submissions are made via the CDRH Customer Collaboration Portal (CDRH Portal) using the eSTAR electronic template.
2. Risk Classification
The FDA uses a three-tier risk classification:
| Class | Risk | Regulatory controls | Typical pathway | Examples |
|---|---|---|---|---|
| Class I | Low | General controls | Mostly exempt; registration + listing only | Examination gloves, tongue depressors, bandages |
| Class II | Moderate | General + special controls | 510(k) | Infusion pumps, X-ray systems, condoms |
| Class III | High | General controls + premarket approval | PMA | Pacemakers, implantable defibrillators, artificial heart valves |
The FDA has classified about 1,700 generic device types across 16 medical specialty panels. About 99% of devices are Class I or Class II (“moderate risk”).
3. Four Main Pathways to Market
1. 510(k) Premarket Notification
Applies to: Class II and a small number of Class I/III devices
Core requirement: Demonstrate Substantial Equivalence (SE) to a legally marketed “predicate device” –
-
Same intended use, or
-
Different technological characteristics that do not raise new questions of safety and effectiveness
Subtypes: Traditional, Abbreviated (based on recognized standards), Special (for minor modifications to the manufacturer’s own device)
FDA review goal (MDUFA V): 95% within 90 FDA days (FY2025–2027). Actual end-to-end time: typically 6–9 months (including interaction with the FDA)
2. De Novo Classification Request
Applies to: novel low-to-moderate risk devices with no legally marketed predicate
Mechanism: Avoids automatic Class III designation requiring a PMA. A De Novo grant creates a new Class I/II classification, so future similar products can use the 510(k) pathway.
Important change: From 1 October 2025, all De Novo requests must be submitted using the eSTAR electronic template.
FDA review goal: 70% within 150 FDA days (FY2025–2027)
3. PMA (Premarket Approval)
Applies to: Class III high-risk devices – the most stringent and comprehensive FDA approval pathway
Core requirements:
-
Sufficient valid scientific evidence of safety and effectiveness
-
Usually includes a pivotal clinical trial
-
Detailed design, manufacturing and QMS information
-
May require review by an Advisory Panel
Review time: 180 statutory days plus interaction time; average total time in FY2025–2027 about 285 days
4. HDE / HUD (Humanitarian Device Exemption)
Applies to: devices for rare diseases (affecting ≤8,000 people per year). A simplified alternative to PMA requiring evidence of “probable benefit” rather than effectiveness.
5. Breakthrough Devices Program
For innovative devices treating serious or life-threatening conditions, offering priority review + expedited communication. More than 100 breakthrough devices had been authorized as of 2025.
4. Requirements for All Manufacturers: Registration & Listing
Whatever the pathway, all manufacturers, initial importers, reprocessors, etc. must complete the following every year:
1. Establishment Registration
-
Register the establishment in the FDA system
-
Foreign manufacturers must designate a U.S. Agent
2. Device Listing
-
List the proprietary name, product code, regulatory status and activities at each establishment
3. Pay the annual fee – FY2026 (1 October 2025 – 30 September 2026): USD 11,423
Key deadlines:
Registration window: 1 October – 31 December each year
Missing it means the establishment is no longer “legally registered”, new submissions are put on hold, and imported products are considered “misbranded”
5. FY2026 User Fees (USD)
| Application type | Standard fee | Small business fee |
|---|---|---|
| Establishment registration (annual) | $11,423 | No reduction |
| 510(k) | $25,072 | $6,517 |
| 513(g) (request for information) | $7,522 | $3,761 |
| De Novo | $167,140 | $41,785 |
| PMA / PDP / PMR / BLA | $556,758 | $139,190 |
| Panel-track supplement | $417,569 | $104,392 |
| 180-day supplement | $83,514 | $20,879 |
| 30-Day Notice | $8,908 | $4,454 |
| Class III annual report | $19,487 | $4,872 |
Small business qualification (SBD Program):
Gross receipts of the company and affiliates ≤ USD 100 M: eligible for reduced fees
Gross receipts ≤ USD 30 M: first PMA / PDP / PMR / BLA fee fully waived
Gross receipts ≤ USD 1 M with demonstrated financial hardship (e.g. bankruptcy): may apply for a registration fee waiver (new in FY2026)
6. A Must for Foreign Manufacturers: the U.S. Agent
Foreign manufacturers must designate a U.S. Agent, whose responsibilities include:
-
Receiving FDA inspection notices and urgent communications on the manufacturer’s behalf
-
Helping to arrange FDA inspections
-
Not taking over the manufacturer’s product compliance responsibilities
-
Each foreign establishment may designate only one U.S. Agent
The U.S. Agent must be a legal entity or individual located in the US and reachable by the FDA during business hours.
7. Registration Process (510(k) Example)
Step 1 — Product classification: Search the FDA Product Classification Database. If unsure, submit a 513(g) Request (USD 7,522) for official feedback.
Step 2 — Select a predicate device: Find a legally marketed device in the 510(k) database that is substantially equivalent to your product.
Step 3 — Establish a QMS (21 CFR 820, the QSR)
-
Important update: The FDA has issued the QMSR (amended 21 CFR 820), aligning 21 CFR 820 with ISO 13485:2016
-
Effective date: 2 February 2026 – FDA QMS inspections will then follow the ISO 13485 framework
-
A historic alignment of the FDA with the global QMS standard
Step 4 — Testing and verification
-
Biocompatibility (ISO 10993)
-
Electrical safety (IEC 60601)
-
Software (IEC 62304 + cybersecurity)
-
Sterilization validation
-
Performance testing
Step 5 — Prepare the 510(k) submission using the eSTAR template, including:
-
Device description and intended use
-
Substantial equivalence discussion versus the predicate
-
Risk management (ISO 14971)
-
Performance data
-
Draft labeling
-
Clinical data (if applicable)
Step 6 — Submit via the CDRH Portal + pay the fee
Step 7 — FDA review
-
Acceptance review (first 15 days)
-
Substantive review – you may receive an Additional Information (AI) Request, which must be answered within 180 days
Step 8 — Obtain 510(k) clearance (K number): Once a K number is issued, the device may be legally marketed in the US.
Step 9 — Complete establishment registration and device listing before marketing
8. QMSR Transition: the Key Date of 2 February 2026
The QMSR (Quality Management System Regulation) replaces the current QSR (21 CFR 820). Key changes:
-
Full alignment with ISO 13485:2016 – no need to maintain two separate systems
-
Certain FDA-specific requirements are retained (e.g. 21 CFR 803 adverse event reporting, traceability)
-
Change in inspection approach: FDA investigators will take a perspective similar to that of ISO 13485 auditors
Strategic significance: The compliance threshold for foreign manufacturers already holding ISO 13485 certification to enter the US market is significantly lowered, and MDSAP certification becomes even more valuable.
9. Post-Market Obligations
-
Medical Device Reporting (MDR) – under 21 CFR 803, manufacturers must report within 30 days (serious injury / death / malfunction), and in certain cases within 5 working days
-
Corrections and removals reporting – under 21 CFR 806, recalls must be reported in writing within 10 working days
-
UDI (Unique Device Identification) – phased implementation by risk class has been completed
-
MDSAP – MDSAP audit reports can replace some routine FDA inspections
-
Annual report (Class III) – USD 19,487 (FY2026)
-
Import entry and customs clearance – via the Import Trade Auxiliary Communication System (ITACS)
10. IVDs and Special Product Categories
IVDs:
-
Most follow the 510(k) or PMA pathway
-
CLIA Waiver applications – for simple-to-use tests (home use, point of care)
-
LDTs (laboratory developed tests): the FDA’s April 2024 final rule brings LDTs under device regulation, phased in through 2028
SaMD / AI medical software:
-
Predetermined Change Control Plans (PCCP) allow pre-authorized model update mechanisms
-
The AI/ML-Based Software Action Plan continues to guide regulation of AI devices
Combination products:
-
Coordinated by the Office of Combination Products (OCP)
-
The lead center (CDRH / CDER / CBER) is determined by the primary mode of action (PMOA)
11. Recent Key Changes (Timeline)
| Date | Event |
|---|---|
| April 2024 | FDA final rule brings LDTs under device regulation |
| 31 July 2025 | FY2026 user fees published; registration fee rises to USD 11,423 (+23%) |
| 1 October 2025 | eSTAR electronic submission mandatory for De Novo requests |
| 1 October 2025 | FY2026 user fees take effect |
| 2 February 2026 | QMSR replaces QSR, aligned with ISO 13485:2016 |
| FY2025–2027 | MDUFA V continues, with review goals for 510(k)/De Novo/PMA |
| 2028 | LDT regulation fully implemented |
Strategic Value and Challenges
Opportunities:
-
The world’s largest market – about 40%+ of global medical device sales
-
The most widely recognized approval – FDA approval underpins expedited pathways in almost every country
-
The 510(k) pathway is relatively mature and predictable – 95% completed within 90 FDA days
-
Small business fee reductions – gross receipts ≤ USD 100M can cut fees substantially
-
MDSAP certification can reduce routine FDA inspections
-
QMSR aligned with ISO 13485 – reduces global compliance complexity
-
Breakthrough Devices and PCCP – friendly to innovative devices
Challenges:
-
Application fees are among the highest in the world (a PMA costs nearly USD 557,000)
-
Demanding testing and verification requirements
-
Increasingly rigorous substantial equivalence arguments for 510(k)
-
Short adverse event reporting deadlines (30 days / 5 working days)
-
Complex import and customs clearance procedures
-
Mandatory U.S. Agent, plus time-zone and cultural communication challenges
-
The new LDT rule brings major changes to the IVD industry
-
Cybersecurity requirements (Section 524B, 2023) continue to tighten
How EVERBIZ Can Help
EVERBIZ has offices in Shanghai, Hong Kong, Germany, Austria, Singapore, Malaysia, Thailand and Vietnam. For the US market, we offer:
- Product classification and registration pathway planning
- Preparation and review of 510(k), De Novo and PMA submissions
- Local authorized representative / registration holder arrangements
- Submission, responses to deficiency letters and progress tracking
- Post-market compliance maintenance (changes, renewals, adverse event reporting)
Contact us for a US medical device registration plan and quote, or see our registration services for other countries and regions.
US Medical Device Registration FAQs
Which agency regulates medical devices in the US?
The Center for Devices and Radiological Health (CDRH) of the US Food and Drug Administration (FDA).
What are the main pathways to the US market?
Depending on risk class, the main pathways are exemption, 510(k) premarket notification, De Novo classification request and PMA premarket approval. Most Class II products follow the 510(k) pathway.
Do foreign manufacturers need a U.S. Agent?
Yes. Foreign manufacturers must designate a U.S. Agent, complete Establishment Registration and Device Listing, and pay the annual registration fee.
When does the QMSR take effect?
The FDA Quality Management System Regulation (QMSR) takes effect on 2 February 2026. Based on ISO 13485:2016, it replaces the former QSR (21 CFR 820).
Related Reading
- Major Update to the FDA Recognized Consensus Standards Database
- Philippine FDA Launches the eSumbong Complaint System: What Medical Device Companies Need to Know
- Market Access Guide for Exporting Veterinary X-Ray Machines to the US
- 2026 Guide to Saudi SFDA Medical Device Registration and Doing Business: New Regulations and Market Access Explained
- FDA: The Complete Compliance Path for Medical Device Biocompatibility Submissions