1. Basic Concepts
Biocompatibility: As defined in ISO 10993-1, the ability of a medical device material to perform with an appropriate host response in a specific application – i.e. the device performs its intended function without causing unacceptable local or systemic adverse effects in the recipient.
Biological evaluation: a systematic approach to assessing the biological risks that may arise when a medical device contacts the human body, ensuring product safety.
Core purposes:
-
Assess the safety of materials in contact with the human body
-
Identify potential biological risks (toxicity, sensitization, carcinogenicity, etc.)
-
Provide the biological basis for product registration
-
Meet global regulatory requirements
Important change: modern biological evaluation has moved from “run every test” to an integrated approach of “risk assessment + chemical characterization + necessary testing”.
2. Key Regulations and Standards
ISO 10993 series (international)
| Standard | Content |
|---|---|
| ISO 10993-1:2018 | Evaluation and testing framework (overarching, core) |
| ISO 10993-2 | Animal welfare requirements |
| ISO 10993-3:2014 | Tests for genotoxicity, carcinogenicity and reproductive toxicity |
| ISO 10993-4:2017 | Selection of tests for interactions with blood |
| ISO 10993-5:2009 | Tests for in vitro cytotoxicity |
| ISO 10993-6:2016 | Tests for local effects after implantation |
| ISO 10993-7:2008+A1:2019 | Ethylene oxide sterilization residuals |
| ISO 10993-9:2019 | Framework for identification and quantification of degradation products |
| ISO 10993-10:2021 | Tests for skin sensitization (irritation moved to Part 23) |
| ISO 10993-11:2017 | Tests for systemic toxicity |
| ISO 10993-12:2021 | Sample preparation and reference materials |
| ISO 10993-13:2010 | Degradation products from polymeric devices |
| ISO 10993-14:2001 | Degradation products from ceramics |
| ISO 10993-15:2019 | Degradation products from metals and alloys |
| ISO 10993-16:2017 | Toxicokinetic study design for degradation products and leachables |
| ISO 10993-17:2023 | Toxicological risk assessment of medical device constituents |
| ISO 10993-18:2020+A1:2022 | Chemical characterization |
| ISO 10993-22:2017 | Nanomaterials |
| ISO 10993-23:2021 | Tests for irritation (split from Part 10) |
FDA
| Document | Content |
|---|---|
| FDA Guidance (2020) | “Use of International Standard ISO 10993-1, Biological evaluation of medical devices” |
| FDA Draft Guidance (2021) | Chemical analysis for biocompatibility assessment |
China
| Standard | Content |
|---|---|
| GB/T 16886 series | Identical adoption of the ISO 10993 series |
| Guidelines for Biological Evaluation and Review of Medical Devices | NMPA technical guidance |
European Union
-
MDR (EU) 2017/745 emphasizes life-cycle biological evaluation
3. Key Terms
| Term | Meaning |
|---|---|
| Nature of contact | Surface device / externally communicating device / implant device |
| Duration of contact | Limited (≤24 h), prolonged (>24 h to 30 d), long-term (>30 d) |
| Contact site | Skin, mucosa, breached surfaces, blood, bone/tissue, circulatory system, etc. |
| Biological endpoints | Specific evaluation items such as cytotoxicity, sensitization, irritation and systemic toxicity |
| Biological evaluation plan (BEP) | Biological Evaluation Plan |
| Biological evaluation report (BER) | Biological Evaluation Report |
| Chemical characterization | Identifying material composition and released substances |
| Toxicological risk assessment (TRA) | Toxicological Risk Assessment |
4. Device Categorization (ISO 10993-1)
Biological evaluation endpoints are determined by the device’s nature of contact + duration of contact.
By nature of contact
| Category | Subcategory | Examples |
|---|---|---|
| Surface devices | Intact skin | Stethoscopes, ECG electrodes |
| Mucosal membranes | Contact lenses, endoscopes | |
| Breached/compromised surfaces | Wound dressings, burn dressings | |
| Externally communicating devices | Blood path, indirect | Infusion sets, blood collection sets |
| Tissue/bone/dentin | Dental fillings, bone cement | |
| Circulating blood | Hemodialyzers, intravascular catheters | |
| Implant devices | Tissue/bone | Bone screws, artificial joints, sutures |
| Blood | Pacemakers, artificial heart valves |
By duration of contact
| Duration | Range |
|---|---|
| A (limited) | ≤24 hours |
| B (prolonged) | >24 hours to 30 days |
| C (long-term) | >30 days |
5. ISO 10993-1 Evaluation Matrix (2018 edition)
The table below is a simplified biological evaluation endpoint matrix; in practice, the tests are determined by consulting this table:
Surface devices
| Contact site | Contact duration | Cytotoxicity | Sensitization | Irritation | Acute systemic toxicity | Subacute/subchronic toxicity | Genotoxicity | Implantation | Chronic toxicity | Carcinogenicity | Reproductive toxicity | Degradation | Hemocompatibility |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Intact skin A/B/C | ● | ● | ● | ||||||||||
| Mucosa A | ● | ● | ● | ||||||||||
| Mucosa B/C | ● | ● | ● | ○ | ○ | ○ | |||||||
| Breached surface A/B/C | ● | ● | ● | ○ | ○ | ○ |
Externally communicating devices
| Contact site | Contact duration | Cytotoxicity | Sensitization | Irritation | Acute systemic toxicity | Subacute/subchronic toxicity | Genotoxicity | Implantation | Chronic toxicity | Carcinogenicity | Reproductive toxicity | Degradation | Hemocompatibility |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Indirect blood path A | ● | ● | ● | ● | ● | ||||||||
| Indirect blood path B/C | ● | ● | ● | ● | ● | ● | ○ | ○ | ● | ||||
| Tissue/bone B/C | ● | ● | ● | ● | ● | ● | ● | ○ | ○ | ○ | |||
| Circulating blood A | ● | ● | ● | ● | ○ | ● | |||||||
| Circulating blood B/C | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
Implant devices
| Contact site | Contact duration | Cytotoxicity | Sensitization | Irritation | Acute systemic toxicity | Subacute/subchronic toxicity | Genotoxicity | Implantation | Chronic toxicity | Carcinogenicity | Reproductive toxicity | Degradation | Hemocompatibility |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Tissue/bone A | ● | ● | ● | ● | ● | ||||||||
| Tissue/bone B/C | ● | ● | ● | ● | ● | ● | ● | ● | ● | ○ | |||
| Blood A | ● | ● | ● | ● | ○ | ● | ● | ||||||
| Blood B/C | ● | ● | ● | ● | ● | ● | ● | ● | ● | ● |
● = usually needs to be evaluated; ○ = evaluate case by case
Consult the full ISO 10993-1 matrix for actual projects
6. Key Biological Endpoints in Detail
1. Cytotoxicity – ISO 10993-5
Purpose: assess the toxicity of materials to cells in vitro. Common methods:
-
MTT assay: measures cellular metabolic activity (most common)
-
XTT and MTS assays: similar to MTT
-
Neutral red uptake
-
Colony formation assay
Test approaches:
-
Extract method: cells treated with an extract
-
Direct contact method: material in direct contact with cells
-
Indirect contact method: agar overlay
Acceptance criteria:
-
Cell viability ≥70%: non-cytotoxic
-
50–70%: mild
-
30–50%: moderate
2. Sensitization – ISO 10993-10
Purpose: assess whether the material causes delayed-type hypersensitivity. Common methods:
-
Guinea pig maximization test (GPMT): the classic method
-
Buehler test: closed patch
-
Murine local lymph node assay (LLNA): better for animal welfare, preferred
Evaluation: sensitization rate, skin reaction scores
3. Irritation and intracutaneous reactivity – ISO 10993-23
Purpose: assess irritation of the skin, mucosa, eyes, etc. Common methods:
-
Rabbit intracutaneous reactivity test: the classic method
-
Skin irritation test
-
Eye irritation test
-
Oral/vaginal/rectal mucosal irritation tests
-
Reconstructed human epidermis (RhE) model: an in vitro alternative (increasingly accepted)
4. Acute systemic toxicity – ISO 10993-11
Purpose: toxic effects within 24–72 hours after a single exposure. Method: extracts administered to mice/rats intravenously, intraperitoneally, orally or subcutaneously. Observations: clinical signs, body weight, mortality
5. Subacute/subchronic/chronic toxicity – ISO 10993-11
| Type | Contact duration |
|---|---|
| Subacute | 14–28 days |
| Subchronic | 90 days |
| Chronic | ≥6 months |
Observations: body weight, hematology, clinical chemistry, urinalysis, histopathology
6. Genotoxicity – ISO 10993-3
Purpose: assess DNA damage and gene mutations. Common battery (three tests):
-
Ames test (bacterial reverse mutation): gene mutations
-
Chromosomal aberration test or micronucleus test: chromosomal damage
-
Mouse lymphoma assay or in vivo micronucleus test: in vivo confirmation
7. Implantation – ISO 10993-6
Purpose: assess local tissue reactions after implantation. Sites: subcutaneous, muscle, bone. Time points: 1, 4, 12, 26, 52 weeks, etc. Evaluation: inflammation, fibrous encapsulation, necrosis, vascularization, tissue reaction scores
8. Hemocompatibility – ISO 10993-4
Purpose: assess reactions when the material contacts blood. Evaluation categories:
-
Hemolysis (routinely required)
-
Coagulation (PT, APTT, TT)
-
Platelets (count, activation, aggregation)
-
Hematology (blood cell counts)
-
Complement system (C3a, C5a, SC5b-9)
Note: a hemolysis test is required for almost all blood-contacting devices.
9. Carcinogenicity – ISO 10993-3
Purpose: assess whether long-term exposure causes cancer. Duration: ≥2 years (lifetime). Alternative: may be waived if the material’s chemical composition is well defined and toxicological data are sufficient
10. Reproductive and developmental toxicity – ISO 10993-3
Purpose: assess effects on the reproductive system, embryonic development and offspring. Applies to: devices that may contact reproductive organs, pregnant women or newborns
11. Degradation products – ISO 10993-9/13/14/15
Purpose: assess the products of material degradation and their toxicity. Categories:
-
Polymers (Part 13): hydrolysis and oxidation products
-
Ceramics (Part 14): ion release
-
Metals (Part 15): metal ion release
-
Bioabsorbable materials: special attention
12. Chemical characterization – ISO 10993-18
See our Chemical Characterization page – the core input to modern biological evaluation.
7. Biological Evaluation Process (ISO 10993-1)
1. Gather device information (materials, processes, sterilization, clinical use) ↓ 2. Risk management (ISO 14971) ↓ 3. Chemical characterization (ISO 10993-18) + literature review ↓ 4. Biological evaluation plan (BEP) ↓ 5. Gap analysis Are existing data sufficient? ↓ ┌────┴────┐ ↓ ↓ Sufficient data Insufficient data ↓ ↓ │ Perform necessary biological tests ↓ ↓ └────┬────┘ ↓ 6. Toxicological risk assessment (ISO 10993-17) ↓ 7. Biological evaluation report (BER) ↓ 8. Ongoing evaluation (life cycle)
8. The Modern Approach to Biological Evaluation
Traditional vs. modern approach
| Dimension | Traditional approach | Modern approach (2018 edition onwards) |
|---|---|---|
| Starting point | Run every test in the matrix | Based on risk assessment |
| Chemical information | Secondary | Core input |
| Animal testing | Extensive | Minimized (3R principles) |
| Use of literature | Limited | Make full use of existing data |
| Depth of evaluation | Test-result driven | Risk driven |
Reducing animal testing (3R principles)
-
Replacement: in vitro tests, in silico models
-
Reduction: optimized design, fewer animals
-
Refinement: improved animal welfare
9. Sample Preparation (ISO 10993-12)
Extraction conditions
| Contact duration | Recommended extraction conditions |
|---|---|
| ≤24 hours | 37 ±1 °C, 72 h or 50 ±2 °C, 72 h |
| 24h~30d | 37 ±1 °C, 72 h or 70 ±2 °C, 24 h |
| >30d | 37 ±1 °C, 72 h or 50 ±2 °C, 72 h |
| Exaggerated | 121 °C, 1 hour (heat-resistant materials only) |
Extraction vehicles
-
Polar: saline, culture medium, water for injection
-
Non-polar: vegetable oil, polyethylene glycol
-
Where necessary: ethanol/water, DMSO
Extraction ratio (surface area/volume)
| Material thickness/shape | Ratio |
|---|---|
| Thickness <0.5 mm | 6 cm²/mL |
| Thickness 0.5–1 mm | 3 cm²/mL |
| Thickness >1 mm | 3 cm²/mL or 6 cm²/mL |
| Irregular/granular | 0.2 g/mL |
10. Typical Evaluation Endpoints for Different Devices
| Device type | Typical endpoints |
|---|---|
| Adhesive bandages/dressings | Cytotoxicity, sensitization, irritation |
| Contact lenses | Cytotoxicity, sensitization, eye irritation |
| Infusion sets | Cytotoxicity, sensitization, irritation, acute systemic toxicity, hemolysis |
| Intravenous catheters (short-term) | Cytotoxicity, sensitization, irritation, acute systemic toxicity, hemolysis, chemical characterization |
| Coronary stents | Full set (except a few special items) |
| Artificial joints | Cytotoxicity, sensitization, irritation, systemic toxicity, genotoxicity, implantation, degradation, chemical characterization |
| Bone cement | Cytotoxicity, sensitization, irritation, acute systemic toxicity, implantation, degradation |
| Surgical instruments (reusable) | Cytotoxicity, sensitization, irritation |
| Dental implants | Cytotoxicity, sensitization, irritation, implantation, genotoxicity, degradation |
11. Factors Affecting Biological Evaluation
1. Material factors
-
Chemical composition, additives
-
Surface condition (coatings, roughness)
-
Physical form (sheet, granules, fibers)
2. Process factors
-
Residual processing aids (mold release agents, lubricants)
-
Cleaning processes
-
Sterilization method (EO, irradiation, moist heat)
-
Sterilization residues (EO, ECH)
3. Clinical factors
-
Contact site
-
Contact time and frequency
-
Patient population (children, pregnant women, immunocompromised)
-
Cumulative exposure
4. Packaging factors
-
Leachables from packaging materials
-
Effect of storage conditions on materials
12. Common Challenges and Pitfalls
1. Misunderstanding the matrix
-
Misjudging the nature or duration of contact, leading to missing or excessive endpoints
2. Improper sample preparation
-
Not reflecting actual processing (after sterilization, cleaning)
-
Unscientific extraction conditions
3. Insufficient chemical characterization
-
TOC alone cannot replace full chemical characterization
-
Incorrect AET calculation
4. Weak literature review
-
Not making full use of data on marketed equivalent devices
-
Insufficient chemical equivalence arguments
5. Misinterpreting test results
-
Giving up as soon as mild cytotoxicity is found
-
Not considering the clinical context as a whole
6. Lack of change management
-
Material, process or supplier changes not re-evaluated
7. Regulatory differences
-
The FDA and NMPA have different requirements for some tests
-
The EU MDR is stricter on post-market surveillance
13. Regulatory Considerations
FDA
-
Accepts ISO 10993-1 as the main basis
-
Emphasizes chemical characterization + toxicological assessment to reduce animal testing
-
Requires complete BEP/BER submissions
-
In some cases requires chemical characterization as evidence of biocompatibility equivalence
EU MDR
-
Strict life-cycle evaluation
-
Ongoing post-market surveillance requirements
-
Emphasis on integrating risk management and clinical evaluation
China NMPA
-
Identical adoption of the ISO 10993 series (GB/T 16886)
-
Accepts comparison with equivalent devices to waive some tests
-
Stricter requirements for novel materials
-
Emphasis on traditional animal test data (changing)
International trends
-
Global adoption of the 3R principles
-
Chemical characterization + TRA becoming mainstream
-
In vitro alternatives increasingly accepted (e.g. RhE models)
14. Key Contents of a Biological Evaluation Report (BER)
-
Device information: name, intended use, composition, clinical application
-
Material information: formulation, suppliers, processing
-
Contact category: nature, duration and site of contact
-
Risk management input: identified biological risks
-
Chemical characterization results
-
Review of existing data: literature, equivalent devices, historical data
-
Gap analysis: determining which tests need to be added
-
Summary of test reports: method, results and conclusion for each test
-
Toxicological risk assessment (TRA)
-
Overall evaluation and conclusion: whether the risk is acceptable
-
Life-cycle monitoring plan
15. Typical Project Timelines and Costs
Timelines for a complete biocompatibility evaluation
| Endpoint | Timeline |
|---|---|
| Cytotoxicity | 2–4 weeks |
| Sensitization (GPMT/LLNA) | 6–8 weeks |
| Irritation (intracutaneous/skin) | 2–4 weeks |
| Acute systemic toxicity | 4–6 weeks |
| Subacute/subchronic toxicity | 4–6 months |
| Genotoxicity (three-test battery) | 2–4 months |
| Implantation | Depending on duration, 3–12 months |
| Chronic toxicity/carcinogenicity | 1–2 years |
| Chemical characterization + TRA | 2–4 months |
Cost ranges
-
Basic package (cytotoxicity + sensitization + irritation): tens of thousands of RMB
-
Intermediate package (including systemic toxicity, implantation): from over a hundred thousand to several hundred thousand RMB
-
Complete package (high-risk implants): from several hundred thousand to a million RMB
-
Chemical characterization + TRA: from tens of thousands to hundreds of thousands of RMB
16. Recommendations and Best Practices
1. Plan early
-
Start planning the biological evaluation early in product development
-
Consider biocompatibility when selecting materials
2. Prioritize chemical characterization
-
Chemical data can guide the design of subsequent tests
-
May waive a large number of animal tests
3. Make full use of historical data
-
Equivalent device comparisons, literature reviews, supplier data
-
Justify the applicability of the data
4. Choosing a third-party laboratory
-
Choose GLP-qualified laboratories
-
China: CMA, CNAS accreditation
-
International: FDA GLP, OECD GLP
5. Life-cycle perspective
-
Not just a one-time evaluation at registration
-
Ongoing for changes and post-market surveillance
6. Interdisciplinary collaboration
-
Collaboration across materials, toxicology, clinical and regulatory
-
Engage toxicology experts where necessary
7. Complete documentation
-
Write the BEP/BER using structured templates
-
Traceable data and well-founded decisions
17. Emerging Trends
1. Microphysiological systems (organ-on-chip)
-
Simulating organ function in vitro, with potential to replace animal testing
2. Computational toxicology (in silico)
-
Predicting toxicity with QSAR and PBPK models
3. High-throughput screening
-
Rapid early screening of materials
4. Digital twins
-
Simulating the long-term behavior of devices in the body
5. Evaluation of nanomaterials
-
The new ISO 10993-22 specifically addresses nanomaterials
6. 3D-printed/personalized devices
-
Process diversity brings new challenges
7. Bioabsorbable materials
-
Dynamic evaluation of the degradation process
18. Relationship Between Biocompatibility and Other Tests
Material selection → chemical characterization → biocompatibility evaluation ↓ ↓ Toxicological assessment Clinical evaluation ↓ ↓ Risk management ←——————————┘ ↓ Registration submission + post-market surveillance
-
With chemical characterization: mutual inputs; chemical data guide the biological evaluation
-
With the sterile barrier: the sterilization method affects biocompatibility
-
With aging tests: material degradation after aging may affect biocompatibility
-
With clinical evaluation: biological data support clinical safety
-
With risk management: biological risk is part of the overall risk